VPH Gratings
N.B. Both 740 l/mm and 3300 l/mm VPHGs have more strongly-peaked blaze functions than the SRGs, however, the VPHG diffraction efficiencies are still greater than the SRG efficiency at the edge of the Bench band-pass. 740 l/mm The figure below shows a direct comparison of the 740 l/mm VPH grating with the 790 l/mm surface ruled grating from the KPNO Mayall 4m RC psectrograph collection. The two gratings were both used with the Bench spectrograph to obtain dome flats and comparison lamp data. The plots show the flux ratios of the 740 l/mm VPH to the 90 l/mm surface ruled grating in orders 1 and 2. For each of the wavelngth regions measured, the VPH grating has to be rotated to the optimum angle of incidence and diffraction, indicated as 17 °, 19 °, etc. The plots demonstrate that the VPH grating is has much higher throughput, yet gives comparable spectral resolution.
3300 l/mm
The figure below shows predicted spectral resolving power
and wavelength coverage as a function of angle of incidence.
The first plot below shows the measured ratios of flux in
comparison spectra taken with the 316@63.4 echelle and the
3300 l/mm VPH for 5130 Å (&alpha=57.8 degrees) The
second figure shows the ratios of line widths in the direction
of dispersion for several lines. The anamorphic demagnification
of the echelle narrows the lines about 30% compared to the
VPH. The measured &Delta&lambda/&lambda for the VPH from
this configuration was 16,700. On a prectical note, the
setting of the central wavelength is probably no better than
5 Å.
The figure below (from M. Bershady) shows predicted relative efficiencies for the 3300 VPH for several grating angles to demonstrate how the efficiency varies with grating angle. Note that the measured data shown above were taken at a grating angle of 57.8 degrees. N.B. the width of the orders seems to be broader than sugested by these calculations. Additional data will be added as it becomes available.
|
||||||||||||||||||||